Medical Vacuum Drying Cabinet
Medical Vacuum Drying Cabinet Specifications:
Model | UVDC-100 |
---|---|
Capacity | 100L (Double Chamber) |
Internal Size (W×D×H) | 320x220x720mm |
External Size (W×D×H) | 880x530x1665mm |
Door | Vertical manual door, double door (front and back), with glass viewing window |
Display | 7 inch touch screen |
Working Mode | Chamber a mode, chamber B mode, double chamber mode (A and B)(Each mode has 5 fixed time modes and 6 DIY modes) |
Alarm | High temperature alarm of drying chamber; Low temperature alarm of drying chamber; Low pressure alarm of drying chamber; Low evacuation rate alarm; |
Safety Protection | Over voltage protection of power supply; Over temperature protection of drying chamber; Anti-pinch hand function. |
Heating Type | Electric heating |
Drying Temperature | 45~65℃ |
Drying Time | 1~999min, adjustable |
Drying Chamber Leakage Rate | ≤5Pa/s (Leakage rate was determined under the conditions of no-load, constant temperature of drying chamber, test pressure of 3kPa and test time of 1h.) |
Ambient Temperature | 5~35℃ |
Ambient Humidity | Max.relative humidity is 80%, when ambient temperature<31℃; Max.relative humidity is 67%, when ambient temperature is 35℃. |
Air Filter | ≥99.5% efficiency at 0.3µm |
Working Pressure | 3kPa |
Max. Pressure Resistance | 1kPa |
Vacuum Rate | Vacuum time ≤ 1min, from atmospheric pressure to 3kPa |
Material | Internal: Aluminium alloy, thickness is 16mm External: 304 Stainless steel, with a whole piece of soundproofing cotton. |
Tray | Chamber A: 2pcs trays Chamber B: 2pcs trays Total 4pcs, 304 stainless steel |
Standard | Printer, trays, temperature sensor |
Consumption | 3.3kW |
Power Supply | AC220V, 50Hz |
Shipping Size(W×D×H) | 1070x680x1820mm |
Gross Weight | 318kg |
Medical Vacuum Drying Cabinet Features:
Medical vacuum drying cabinet is suitable for drying various types of endoscopes, biopsy forceps, precision instruments, complex instruments, instruments that are not resistant to high temperatures, and routine instruments used in operating rooms, disinfection supply centers, and endoscopy centers in medical institutions.
1. With 7 inch touch screen
Display temperature, Pressure, Remaining drying time, Drying stage.
2. Front and back double doors
Double doors interlock, when front door is opened, the back door can not be opened, effectively isolating the contaminated area and clean area, good sealing.
3. Glass observation window
Hollow double-layer tempered glass, with LED lights, illuminate the drying chamber, easy to observe whether the drying is completed.
4. Air replacement
Automatic ventilation, ventilation times and intervals can be set freely.
5. Data Traceability
Industrial switches with traceability system grid interface.
6. Working mode
Divided into Chamber A mode, Chamber B mode, Double Chamber mode. Each mode can run separately, can also run at the same time.
Each working mode is divided into 5 fixed time modes and 6 custom modes.
It can be set freely according to the drying items, suitable for different types of items drying.
7. Standard printer
It can record the operating information of the equipment, print various parameters during the drying process, including temperature, time, pressure, etc., as well as fault alarm information printing, saving time is long, easy to view and analyze.
8. Pressure transmitter
Built-in pressure transmitter automatically detects the pressure in the drying chamber to confirm that the pressure in the drying chamber is accurate and effective.
9. Temperature sensor
Standard with external temperature sensor, automatically detects the temperature of the drying bulkhead to ensure the accurate and effective temperature of the drying chamber.
10. Safety protection
Including over-voltage protection of power supply, over-temperature