PCR Cabinet PCR-800
PCR Cabinet Specifications:
Model | UPCR-800 | |
---|---|---|
External Size (WxDxH)(mm) | 710 x 680 x 1050 (Without Base Stand) 710 x 680 x 1680 ( With Base Stand) | |
Internal Size (WxDxH)(mm) | 610x590x625 | |
Work Surface Height | 750mm | |
Control System | Display | Microprocessor controller with LCD display, display the operating status of each function |
Interlock | Front sliding sash interlock with fan and UV lamp | |
Life Indicator | HEPA filter and UV life indicators facilitate a timely replacement plan | |
Airflow System | Airflow Velocity | 0.2~0.5m/s, Speed adjustable |
Air filtration System | Pre-Filter | 1pcs, Polyester fiber,washable |
HEPA Filter | 1 pcs, Efficiency: > 99.995% at particle size of 0.3μm | |
Cleanliness Grade | ISO Class 5 (formerly Class 100) | |
Partial air circulation | Perforations at the back wall of the work zone which are designed to eliminate air turbulence and the possibility of dead air corners in the work zone. | |
Cabinet Construction | Main Body | Electro-Galvanized steel with antimicrobial powder coating |
Work Surface | 304 Stainless steel, an angled front, rounded work surface front edge | |
Front Window | Manual sliding sash: 5 mm UV resistant tempered glass | |
Electrical System | UV Lamp | 10W×2 |
With timing function from 0 to 999 minutes Interlock with front sliding sash. Emission of 253.7 nanometers for best sterilization effect. | ||
LED Lamp | 8W×2 | |
Consumption | 149W (Not include socket loading power consumption) | |
Power Supply | AC220V±10%, 50/60HZ; 110V±10%, 60Hz | |
Illumination | ≥500Lux | |
Sound Emission | ≤62dB(A) | |
Standard Accessory | 1. LED Lamp×2; 2. UV Lamp×2; 3. Waterproof Socket×1; 4. Shelves with IV Bar | |
Optional Accessory | 1. Base stand; 2. Gas Tap | |
Net Weight | 80kg (Without base stand), 130kg (With base stand) | |
Gross Weight | 115kg (Without base stand), 165kg (With base stand) | |
Shipping Size (W×D×H)(mm) | 920x855x1190 | |
Shipping Volume | 0.94m³ |
PCR Cabinet Features:
- Safety Performance:
- Front sliding sash interlock with fan and UV lamp.
- Front sash and side windows are made of UV-resistant tempered glass.
- The host and sockets have independent capacitors to prevent current overload.
Humanized design:
- Standard shelves with IV bar.
- Memory function in case of power-failure.
- HEPA filter and UV life indicators facilitate a timely replacem ent plan.
- Easy-to-change pre-filter installed could extend the HEPA filter’s life time.
- Equipped with temperature sensor to show the real temperature during operation.
- Waterproof socket provide convenience for small instruments to be used in the operating area.
- Perforations at the back wall of the work zone which are designed to eliminate air turbulence and the possibility of dead air corners in the work zone.